Cleanroom Certification Guide

Cleanroom certification checks that a room or hood meets its ISO class or a standard such as USP <797> or <800>, using tests like particle counts, HEPA leak, airflow and pressure difference. Facilities anywhere in the U.S. can request quotes.

Facilities anywhere in the U.S. can request quotes here.

What certification includes

Typical tests include airborne particle counts to ISO 14644-1, plus tests from ISO 14644-3 such as HEPA filter leak, airflow, pressure difference and airflow visualization. USP <797> certification adds dynamic smoke pattern tests for each hood.

How often to recertify

It depends on the standard. ISO 14644-2 calls for periodic classification annually, extendable with risk assessment and monitoring data. USP <797> requires sterile compounding areas and hoods to be recertified at least every 6 months. CDC's BMBL says biosafety cabinets should be certified at least annually.