ISO 5 Cleanroom Certification
An ISO 5 cleanroom or clean zone allows no more than 3,520 particles 0.5 µm and larger per cubic meter of air under ISO 14644-1. ISO 5 is the class required inside sterile compounding hoods (PECs) under USP <797>. Certification adds HEPA leak, airflow velocity and smoke tests. Compare ISO 5 certification quotes with one request.
Facilities anywhere in the U.S. can request quotes here.
ISO 5 particle limits (ISO 14644-1:2015)
| Particle size | Max particles per m³ |
|---|---|
| ≥0.1 µm | 100,000 |
| ≥0.2 µm | 23,700 |
| ≥0.3 µm | 10,200 |
| ≥0.5 µm | 3,520 |
| ≥1 µm | 832 |
There's no 5 µm limit for ISO 5 in the 2015 edition, because counts of large particles that low aren't statistically reliable. Use the M descriptor if you need to describe large particles. ISO 5 roughly matches "Class 100" in the cancelled US FED-STD-209E.
Where ISO 5 applies
- Compounding hoods: USP <797> PECs (LAFWs, BSCs, CAIs, CACIs) must be certified to ISO Class 5 or better under dynamic conditions, at least every 6 months. USP <825> sets the same for radiopharmaceutical PECs.
- Unidirectional-flow clean zones and rooms specified as ISO 5
What ISO 5 certification includes
- Particle counts at ISO 14644-1 Table A.1 sampling locations
- Airflow velocity and uniformity, HEPA filter leak test, airflow direction/visualization (ISO 14644-3)
- For USP <797> PECs: a dynamic airflow smoke pattern test
What to include in your quote request
- Facility address
- ISO 5 rooms or zones and their area, and the number and type of hoods (LAFW, BSC, CAI, CACI)
- Occupancy state (as-built, at-rest, operational/dynamic)
- Standard (ISO 14644, USP <797>, USP <825>, customer spec)
- Due date
Need quotes? Request quotes with one form →
Are you a certification or testing company? Join as a provider →
How getting quotes works here
- Send one request. Use the quote request form to describe your facility, where it is, the rooms or equipment, and the tests or standard you need.
- We share it with providers. Your request goes to certification and testing companies that cover your area and the type of work.
- Compare and choose. Providers that want the job send quotes. You compare scope and price side by side and pick who to hire, so you don't have to chase each company yourself.
Related
- Laminar flow hood certification
- Biosafety cabinet certification
- USP 797 certification
- ISO 7
- ISO 8
- Smoke studies
- EU GMP Annex 1 qualification
Requests are accepted nationwide. Browse locations.
FAQ
How many particles are allowed in an ISO 5 cleanroom?
Up to 3,520 particles 0.5 µm and larger per cubic meter of air, and 832 at 1 µm, per ISO 14644-1:2015.
Why is there no 5 µm limit for ISO 5?
The 2015 edition removed it because sampling such low counts of large particles isn't statistically reliable. The M descriptor can be used instead.
Is ISO 5 the same as Class 100?
Roughly. Class 100 comes from the cancelled US FED-STD-209E, which ISO 14644-1 replaced.
Does a compounding hood have to be ISO 5?
Yes. USP <797> requires the PEC to be certified to ISO Class 5 or better under dynamic operating conditions.
Is ISO 5 the same as EU GMP Grade A?
They're different systems. EU GMP Annex 1 Grade A also has a limit of 3,520 particles ≥0.5 µm per m³, but it sets its own rules for states, monitoring and microbial limits.
Need quotes? Request quotes with one form →
Are you a certification or testing company? Join as a provider →