Fume Hood Testing and Certification Quotes

Fume hood testing checks that a laboratory chemical fume hood captures and contains vapors. Common tests are face velocity, smoke flow visualization and the ANSI/ASHRAE 110 tracer gas test. OSHA's Laboratory standard requires employers to ensure fume hoods are functioning properly. Send one request to compare quotes from testing providers.

Facilities anywhere in the U.S. can request quotes here.

What fume hood testing is

A chemical fume hood protects the user by pulling air in through the sash opening and exhausting it. Testing confirms it still does that.

ANSI/ASHRAE 110 is the standard method of testing laboratory fume hood performance. It includes:

  • Flow visualization: a smoke challenge to see capture and any escape
  • Face velocity: readings across a grid at the sash opening, averaged
  • Tracer gas test: sulfur hexafluoride (SF6), or a gas of similar molecular weight, released at 4.0 L/min inside the hood, with a detector at a mannequin's breathing zone measuring any escape
  • Sash movement and VAV response tests, for hoods on variable air volume systems

Results can be reported "as manufactured" (AM), "as installed" (AI) or "as used" (AU).

Who needs it

  • Labs that use hazardous chemicals. OSHA's Laboratory standard (29 CFR 1910.1450) requires the Chemical Hygiene Plan to include a requirement that fume hoods are functioning properly, and the specific measures taken to ensure it.
  • New or renovated labs commissioning hoods "as installed"
  • Labs after hood repairs, exhaust fan work or HVAC changes

What to include in your quote request

  • Facility address and lab type (university, pharma, clinical, industrial)
  • Number of hoods, with type (bypass, VAV, low-flow, ductless) and sash type (vertical, horizontal, combination)
  • Make, model and width if known
  • Tests wanted: face velocity and smoke only, or full ASHRAE 110 tracer gas (AM, AI or AU)
  • Required face velocity or control level from your Chemical Hygiene Plan or EHS office
  • Other equipment for the same visit (BSCs, laminar flow hoods)
  • Due date

Need quotes? Request quotes with one form →

Are you a certification or testing company? Join as a provider →

How getting quotes works here

  1. Send one request. Use the quote request form to describe your facility, where it is, the rooms or equipment, and the tests or standard you need.
  2. We share it with providers. Your request goes to certification and testing companies that cover your area and the type of work.
  3. Compare and choose. Providers that want the job send quotes. You compare scope and price side by side and pick who to hire, so you don't have to chase each company yourself.

Related services

Requests are accepted nationwide. Browse locations.

FAQ

What is ASHRAE 110 fume hood testing?

ANSI/ASHRAE 110 is the standard method of testing laboratory fume hood performance. It combines smoke flow visualization, face velocity measurements and a tracer gas containment test using sulfur hexafluoride and a mannequin.

Is fume hood testing required by OSHA?

OSHA's Laboratory standard, 29 CFR 1910.1450, requires a Chemical Hygiene Plan that includes a requirement that fume hoods function properly and the specific measures to ensure it. Your plan sets how and how often hoods are checked.

What is the difference between a fume hood and a biosafety cabinet?

A chemical fume hood protects the user from chemical vapors and exhausts the air. A Class II biosafety cabinet uses HEPA filtration to protect the user, the work and the environment from biological agents, and is certified to NSF/ANSI 49.

What do AM, AI and AU mean in an ASHRAE 110 report?

They describe the test condition: as manufactured (factory test), as installed (in the lab, before use) and as used (in the lab with normal equipment and setup).

Can I get fume hoods and BSCs tested in one visit?

Yes. List every hood and cabinet in one request.