Cleanroom Smoke Study (Airflow Visualization) Quotes
A smoke study makes cleanroom airflow visible so you can confirm it moves in the intended direction. USP <797> requires a dynamic airflow smoke pattern test for each compounding hood at certification and at least every 6 months. ISO 14644-3 calls it the airflow direction test and visualization. Request quotes from multiple providers here.
Facilities anywhere in the U.S. can request quotes here.
What a smoke study is
A tester releases a visible, low-residue fog or smoke and watches, and usually records, how air moves.
- ISO 14644-3, airflow direction test and visualization: shows that airflow direction and velocity uniformity match the design. It can be done at rest to see basic patterns and repeated in operation, simulating actual work.
- USP <797>, dynamic airflow smoke pattern test: done for each PEC under dynamic operating conditions, to show unidirectional airflow and sweeping action over and away from the preparation. Required at certification and at least every 6 months.
- USP <825> requires smoke studies at the critical area of radiopharmaceutical PECs under dynamic conditions.
- EU GMP Annex 1 expects airflow pattern studies at rest and in operation (for example, simulating operator interventions), with video recordings.
Who needs it
- USP <797>, <800> and <825> pharmacies at every certification
- Annex 1 sterile manufacturers qualifying grade A and B areas
- Any cleanroom after layout, equipment or HVAC changes
What to include in your quote request
- Facility address
- Number of hoods (PECs) and rooms to study
- At rest, dynamic/in operation, or both
- Whether you need video recording and a written report
- Standard (USP <797>/<825>, ISO 14644-3, Annex 1)
- Other tests for the same visit
Need quotes? Request quotes with one form →
Are you a certification or testing company? Join as a provider →
How getting quotes works here
- Send one request. Use the quote request form to describe your facility, where it is, the rooms or equipment, and the tests or standard you need.
- We share it with providers. Your request goes to certification and testing companies that cover your area and the type of work.
- Compare and choose. Providers that want the job send quotes. You compare scope and price side by side and pick who to hire, so you don't have to chase each company yourself.
Related
- USP 797 certification
- Laminar flow hood certification
- Airflow testing
- EU GMP Annex 1 qualification
- USP 825
Requests are accepted nationwide. Browse locations.
FAQ
How often does USP 797 require a smoke study?
A dynamic airflow smoke pattern test is required for each PEC at certification and at least every 6 months.
What does "dynamic" mean for a smoke study?
It's done while the hood or room is in normal use, with people and activity, so you see how real work affects airflow.
Is a smoke study the same as airflow testing?
No. Airflow testing measures velocity and volume with instruments. A smoke study shows the airflow's direction and pattern.
Should smoke studies be recorded on video?
EU GMP Annex 1 expects video recordings of airflow visualization. USP <797> requires documented results. Ask for video in your request if you want it.