What's in a cleanroom certification report
A cleanroom certification report should record who tested the room, when and where, the standard and class, the instruments and their calibration, the results at every sampling location, and a statement of compliance or non-compliance. ISO 14644-1 and ISO 14644-3 require that documentation. Results for other scoped tests, such as HEPA leak or airflow, belong in the report too.
Facilities anywhere in the U.S. can request quotes here.
Why one request beats ten forms
On a typical provider's website you fill out a quote form, then repeat that process on the next company's site, and the next. On Certification Quotes you describe the job once—your rooms, hoods, tests and timing—and that request is shared with providers. You compare the quotes that come back and choose who to hire.
Need quotes? Describe the job once, then compare what comes back. Request quotes with one form →
What ISO 14644-1 requires in a classification report
ISO 14644-1:2015, clause 5.4, says the results for each cleanroom or clean zone shall be recorded and submitted as a comprehensive report, with a statement of compliance or non-compliance with the specified air cleanliness class. The report shall include:
- The name and address of the testing organization, and the date the test was performed
- The standard and year: ISO 14644-1:2015
- A clear identification of the room or zone, including adjacent areas if needed, and the coordinates of every sampling location (a diagram can help)
- The designation that was tested: ISO class, occupancy state (as-built, at-rest or operational), and the particle sizes considered
- The test method, any special conditions or departures from the method, and identification of the instrument and its current calibration certificate
- The test results, including particle concentration data for all sampling locations
If macroparticles were quantified, ISO 14644-1 says that information should be included with the report.
Ask for this package in the quote, not only a pass/fail certificate. See ISO 14644 cleanroom certification and ISO 14644 explained.
What ISO 14644-3 requires when other tests are performed
When the visit includes performance tests from ISO 14644-3 (airflow, pressure difference, installed filter leakage, airflow visualization, recovery and others), clause 5 of ISO 14644-3:2019 requires a test report for each test. That report shall include the testing organization's name and address and the test date; a reference to ISO 14644-3:2019; identification of the room or zone and sampling locations; the specified designation (class, occupancy state and particle sizes where relevant); the method, any departures, the apparatus and its current calibration certificate; the results, including data required by the relevant Annex B method, and a compliance statement; plus any other requirements for that specific test.
In plain terms: every test in the scope should come back with method, calibrated instruments, raw or tabulated results, and a pass/fail statement against the agreed criterion. For leak-test methods and acceptance, see HEPA filter testing and DOP vs PAO testing.
USP 797 and biosafety cabinet reports
A USP <797> recertification should document the tests that chapter requires: airflow (velocity, volume, air changes and pressure differential), HEPA filter integrity, total particle counts under dynamic conditions, and a dynamic smoke pattern test for each PEC. If viable air sampling is quoted as a separate item, those results should be reported separately from the non-viable certification. See USP 797 certification.
A Class II BSC field certification should report the NSF/ANSI 49 field tests that were performed, typically inflow and downflow velocity, HEPA filter leak, airflow smoke patterns and site installation checks. See biosafety cabinet certification.
What to ask for when you compare quotes
Put report expectations in the request so every provider is pricing the same deliverable:
- Which standard and class (and occupancy state) the report will certify
- Whether you will receive sampling-location diagrams, raw counts, instrument IDs and calibration certificates
- How failures and retests will be documented
- Report format (PDF, and any form your quality unit, state board or accreditor requires)
- How soon the report will be issued after the visit (ask; this site has no standard turnaround to quote)
Need quotes? Describe the rooms, tests and the report you need to get back. Request quotes with one form →
Related guides and services
FAQ
What must an ISO 14644-1 certification report include?
The testing organization's name and address and the test date; ISO 14644-1:2015; the room and sampling locations; the class, occupancy state and particle sizes; the method, instrument and current calibration certificate; results at every location; and a compliance or non-compliance statement.
Should HEPA leak and airflow tests be in the same report?
Each ISO 14644-3 test must be recorded with method, apparatus, calibration, results and a compliance statement. Those sections may be one combined report or attachments, but they should all come back with the job.
Is a pass/fail certificate enough?
ISO 14644-1 calls for a comprehensive report, not only a pass/fail line. Ask for location data, instrument calibration and the designation that was tested.
What if a test fails?
Ask in the quote how failures, repairs and retests will be documented in the report, and whether the retest is included or priced separately.
Ready to request quotes without filling out a new form for every provider? Request quotes with one form →
Are you a certification or testing company? Join as a provider →
Requests are accepted nationwide. Browse locations.