How often to certify a biosafety cabinet
CDC's BMBL says a Class II biosafety cabinet should be certified when installed, after repair or relocation, and at least annually. Cabinets used as a USP 797 primary engineering control need recertification at least every 6 months. Field certification follows NSF/ANSI 49 and typically includes inflow and downflow velocity, HEPA filter leak and smoke-pattern tests.
Facilities anywhere in the U.S. can request quotes here.
Why one request beats ten forms
On a typical provider's website you fill out a quote form, then repeat that process on the next company's site, and the next. On Certification Quotes you describe the job once—your rooms, hoods, tests and timing—and that request is shared with providers. You compare the quotes that come back and choose who to hire.
Need quotes? Describe the job once, then compare what comes back. Request quotes with one form →
The annual interval, and when it is shorter
CDC and NIH's Biosafety in Microbiological and Biomedical Laboratories (BMBL, 6th edition) says a BSC's operational integrity should be validated before it goes into service and after repair or relocation, and that each BSC should be tested and certified at least annually. Moving a cabinet can damage HEPA filters or seals, which is why a move triggers a new certification.
If the cabinet is used as a PEC for sterile compounding under USP <797>, recertification is required at least every 6 months, and sooner after the cabinet is replaced or relocated. Hazardous drug compounding under USP <800> may use a BSC as a C-PEC; the cabinet still needs field certification, and the room has its own USP <800> requirements.
A customer, biosafety officer or accreditor can also set a shorter interval than the BMBL minimum. Put that requirement in the request.
Get quotes: Biosafety cabinet certification.
What a field certification includes
NSF/ANSI 49 is the U.S. standard for Class II biosafety cabinets (Types A1, A2, B1, B2 and C1). It covers design, construction, performance and field certification. The field-certification annex is a normative part of the standard.
A field certification typically includes:
- Inflow velocity: air entering the front opening, which protects the user
- Downflow velocity: HEPA-filtered air moving down over the work surface, which protects the work
- HEPA filter leak test: supply and exhaust filters
- Airflow smoke patterns: a visual check of containment at the front opening and work area
- Site installation checks: such as alarms and, for connected cabinets, exhaust performance
NSF accredits biosafety cabinet field certifiers, with an enhanced program for certifiers who work in North America. The enhanced program uses written and practical examinations. If you require an NSF-accredited certifier, say so in your request. See also HEPA filter testing for how leak tests are done.
What to include in your quote request
- Facility address and lab type
- Number of cabinets, with make, model, class and type (for example, Class II Type A2) and serial number if handy
- Cabinet width and whether it is canopy-connected or hard-ducted
- Whether any cabinet was recently moved, repaired or had filters replaced
- Whether you need an NSF-accredited certifier, and any report format your biosafety officer or USP <797> program requires
- Other equipment for the same visit (room HEPA filters, laminar flow hoods, fume hoods)
- Due date
Need quotes? List every cabinet once. Request quotes with one form →
Related services and guides
FAQ
How often does a biosafety cabinet need to be certified?
CDC's BMBL says at least annually, plus before first use and after repair or relocation. Cabinets used as USP 797 PECs need recertification at least every 6 months.
What standard is used for BSC field certification?
NSF/ANSI 49, which covers Class II biosafety cabinets, including their field certification.
What tests are done during BSC certification?
Typically inflow velocity, downflow velocity, HEPA filter leak testing, airflow smoke patterns, and site installation checks such as alarms.
Does a BSC need recertification after it is moved?
Yes. BMBL says integrity should be validated after relocation, because moving a cabinet can damage HEPA filters or seals.
Ready to request quotes without filling out a new form for every provider? Request quotes with one form →