HEPA Filter Testing Quotes

HEPA filter leak testing, also called integrity testing, challenges an installed filter with an aerosol and scans the media and seals for leaks. ISO 14644-3 and IEST-RP-CC034 describe the method. USP <797> requires HEPA integrity testing at installation and every 6-month recertification. Send one request to compare quotes.

Facilities anywhere in the U.S. can request quotes here.

What HEPA filter leak testing is

A high-efficiency particulate air (HEPA) filter is designed to capture at least 99.97% of particles 0.3 µm in size. After it's installed, leaks can appear in the media, the filter frame, the gasket or the housing. Leak (integrity) testing finds those leaks so they can be repaired or the filter replaced.

ISO 14644-3 calls this the installed filter system leakage test.

The usual method, described in ISO 14644-3 and IEST-RP-CC034, is an aerosol photometer scan:

  1. An aerosol generator introduces a challenge aerosol (commonly PAO, historically DOP) upstream of the filter
  2. The upstream concentration is measured
  3. A photometer probe scans the downstream face of the filter, the frame and the seals
  4. A leak is typically a reading at or above 0.01% of the upstream challenge

The challenge aerosol is commonly PAO today; DOP was the traditional agent. Learn more: DOP vs PAO testing.

This test is different from particle count classification. Particle counts show whether the room meets its ISO class. Leak testing shows whether a specific filter or seal is passing unfiltered air.

USP <797> requires HEPA filter integrity testing as part of certification and the 6-month recertification of sterile compounding areas and hoods. NSF/ANSI 49 includes HEPA leak testing as part of biosafety cabinet certification.

Who needs it

  • Any cleanroom or clean device with installed HEPA or ULPA filters, at installation and at recertification
  • Sterile compounding pharmacies under USP <797> or USP 800
  • Labs with biosafety cabinets
  • Facilities that have replaced filters or had construction that could have disturbed them

What to include in your quote request

  • Facility address
  • Where the filters are: ceiling grid, terminal housings, hoods or biosafety cabinets
  • Approximate number of HEPA filters (and size if you know it)
  • Whether this is a standalone leak test or part of a full room or hood certification
  • Your standard (ISO 14644-3, USP <797>, NSF/ANSI 49, IEST-RP-CC034)
  • Whether you also want particle counts, airflow or pressure tests on the same visit
  • Last test date and when you need the work done

Need quotes? Request quotes with one form →

Are you a certification or testing company? Join as a provider →

How getting quotes works here

  1. Send one request. Use the quote request form to describe your facility, where it is, the rooms or equipment, and the tests or standard you need.
  2. We share it with providers. Your request goes to certification and testing companies that cover your area and the type of work.
  3. Compare and choose. Providers that want the job send quotes. You compare scope and price side by side and pick who to hire, so you don't have to chase each company yourself.

Related services

Requests are accepted nationwide. Browse locations.

FAQ

What is DOP or PAO testing?

They're names for the same kind of test. DOP (dioctyl phthalate) was the traditional aerosol. PAO (polyalphaolefin) is the common substitute today. A generator introduces the aerosol upstream of the filter, and a photometer scans the downstream face and seals for leaks. Learn more: DOP vs PAO testing.

When should HEPA filters be leak tested?

At installation, after a filter is replaced, and at each recertification. USP <797> includes HEPA integrity testing in its 6-month recertification. After construction or changes that could affect the filters, test again.

What's the leak limit?

IEST-RP-CC034.4 and ISO 14644-3 typically use 0.01% of the upstream challenge as the leak threshold when scanning with a photometer.

Can a leak be repaired?

Sometimes a small leak in the media or a seal can be repaired, following the IEST recommended practice and the filter manufacturer's limits. Larger damage usually means replacing the filter. The certifier reports what they found.

Can I request HEPA testing with a full room certification?

Yes. Add it to the same request as ISO 14644, USP <797> or hood certification so providers can quote one visit.